Design of an ECG front-end considering ST segment distortion - CentraleSupélec
Communication Dans Un Congrès Année : 2023

Design of an ECG front-end considering ST segment distortion

Résumé

ECG ST segment should be monitored for myocardial ischemia diagnosis. But this information needs a very accurate ECG acquisition. This paper presents the design of a chopper amplifier filtering the baseline wander and minimizing the added electronic noise with an acceptable ST segment distortion. A method is proposed to estimate the ST segment distortion and compare it to the International Electrotechnical Commission (IEC) standards. The design is achieved using XFAB 180nm technology and validated at transistor-level simulation for typical and worst cases.
Fichier principal
Vignette du fichier
v_HAL.pdf (713.88 Ko) Télécharger le fichier
Origine Fichiers produits par l'(les) auteur(s)

Dates et versions

hal-04121723 , version 1 (08-06-2023)

Identifiants

Citer

Béatrice Guénégo, Caroline Lelandais-Perrault, Emilie Avignon-Meseldzija, Gérard Sou, Philippe Benabes. Design of an ECG front-end considering ST segment distortion. 2023 IEEE 14th Latin America Symposium on Circuits and Systems (LASCAS), Feb 2023, Quito, Ecuador. pp.1-4, ⟨10.1109/LASCAS56464.2023.10108102⟩. ⟨hal-04121723⟩
65 Consultations
91 Téléchargements

Altmetric

Partager

More